Characterization of micro-SPECT system based on Timepix detector
Keywords: coded aperture, Timepix, SPECT
Topics: Instruments and Methods , Applied Research , Nuclear Medicine
Submitted: 06.05.2025; Published 17.06.2025
Views: 13; Downloaded: 2

How to Cite

V. Rozhkov, I. Hernа́ndez, A. Leyva, A. Zhemchugov. "Characterization of micro-SPECT system based on Timepix detector" Natural Sci. Rev. 2 100303 (2025)
https://doi.org/10.54546/NaturalSciRev.100303
V. Rozhkov1,a, I. Hernа́ndez1,2, A. Leyva3, A. Zhemchugov1
  • 1Joint Institute for Nuclear Research, Dubna, Russia
  • 2Isotopes Center, Mayabeque, Cuba
  • 3Center of Technological Applications and Nuclear Development, Havana, Cuba
  • arozhkov@jinr.ru
Natural Sci. Rev. 2 100303 (2025) 17.06.2025
DOI: 10.54546/NaturalSciRev.100303
PDF

In this work, the characteristics of a prototype SPECT system based on the Timepix readout chip, with a MURA-type encoding mask, were evaluated. The setup has a small FoV and can be used in preclinical studies of drugs on small laboratory animals. Despite many existing test protocols developed and described in pertinent documents of national standard bodies and IAEA recommendations, they are not suitable for microtomographic systems based on semiconductor pixel detectors due to different detector technology, high spatial resolution and small area of interest. To measure their characteristics, special phantoms were developed, with a small “hot region”.

Such micro-SPECT parameters as spatial resolution, contrast, linearity, and system efficiency were studied using 99mTc source. The detector calibration and data preprocessing are described.